Smart Spectral Analysis & QR Traceability System

Smart Color Measurement System, Spectrophotometer Integration, CM-M6 Automation, TEMPOint Engineering, QR Code Traceability, Automated QC Pass/Fail, LabVIEW Instrument Control, Spectral Analysis Solutions

Smart Color Measurement & QR Traceability System

TEMPOint: Experts in Spectroscopy Automation
Automated Spectrophotometer & QR Traceability System

Smart Color Inspection & QR Tracking
(Konica Minolta CM-M6 Automation & QC Integration)

Our Smart Spectral Inspection System seamlessly integrates professional-grade spectrophotometers with TEMPOint’s proprietary automated analysis engine. From “High-Speed Measurement” to “Automated Decision Making” and “QR Labeling,” we provide an end-to-end Smart Manufacturing Solution for paint, plastics, and optical film industries.

⭐ Core Value: Eliminate manual inspection errors with a fully digitized QC workflow and real-time production tracking.


Key Features & Engineering Benefits

  • Automated Pass/Fail Logic: Eliminates human bias by executing logic based on pre-defined tolerances for standardized quality.
  • Multi-Angle Spectral Analysis: Simultaneously captures 25°/45°/75° data for comprehensive color profiles under varying lighting.
  • Instant QR Traceability: Generates and prints QR codes immediately post-test, creating a “Digital Birth Certificate” for every part.
  • Robust Data Handling: Full export of L, a, b, and ΔE metrics in CSV/Excel formats, ready for Big Data and SPC (Statistical Process Control).

📍 Targeted Applications

🚗 Automotive CoatingPrecision monitoring for multi-stage paint lines and finish consistency.
📱 Consumer ElectronicsEnsuring batch-to-batch color uniformity for high-end device housings.
📦 Optical MaterialsPerformance analysis and traceability for functional films and substrates.
Technical Specifications
Measurement Engine KM CM-M6 / Custom Spectrophotometer Integration
Data Metrics L, a, b, ΔL, Δa, Δb, ΔE(76), ΔE(94), CMC, etc.
Control Platform Windows IPC / LabVIEW-based Decision Architecture
Traceability Tagging Auto-generated QR Codes (Includes Status, Date, Lot #)

Optimize Your Smart QC Pipeline

TEMPOint provides Custom Software Development to meet your specific hardware and traceability requirements.

📥 Download Technical Datasheet (PDF)

🔍 Technical Index:

Smart Color Inspection
Spectrophotometer Integration
QR Production History
KM CM-M6 Automation
LabVIEW Development
Industry 4.0 QC

VCSEL measurement system

 

Driven by the demand for mobile phone 3D sensing, automotive LiDAR and optical fiber transmission, the VCSEL component market has gradually grown. Using TEMPOint’s VCSEL LIV measurement system, users can quickly and accurately test the VCSEL performance (single point test, variable current , Variable temperature test).

Application field:

  • Face recognition
  • Gesture Recognition
  • Optical radar LiDAR
  • Time difference ranging ToF
  • Laser cutting

Features:

  • Minimum 0.2nm resolution
  • VCSEL dedicated spectrometer
  • 200 Khz high capture speed
  • Minimum 50us pulse width output
  • The instrument can do absolute value calibration in Taiwan
Measurement data
Peak radiant flux (W)      Average radiant flux (W)
Peak wavelength(nm)    Center wavelength(nm)    Center of gravity wavelength(nm)    Half width(nm)
Radiation voltage(V)     Current(A)     Power(W)     Radiation conversion efficiency
Radiant flux@ time domain        Spectral value
system requirement Suggested demand
Windows 7 32bit / 64bit or higher operating system
1 GB or more RAM
2 GB or more hard disk capacity
Windows 7 32bit / 64bit or higher operating system
4 GB or more RAM
20 GB or more hard disk capacity
System specifications
model Spectrum LoMo For VCSEL
Integrating sphere
material Spectralon
Internal dimensions 5.3”( 13.5 cm )
Reflectivity > 98.5% ( 400 – 1500 nm )
> 95% ( 250 – 2500 nm )
Optical specifications
wavelength 800 – 990nm
Optical resolution 0.2 nm
Pixel interval 0.11 nm
Radiant flux range 10m-100w
Wavelength accuracy ± 0.1 nm
Wavelength reproducibility ± 0.2 nm
Detector capture speed 200 Khz
Detector linearity 99.5%
Detector accuracy ± 0.1%
Trace back NML or NIST
Other specifications
Digital interface USB 2.0
physical dimension 28.5 x 18 x 20 cm3
weight 2 kg
Power Specifications
Rated output 60V / 7.5A for Pulse / 500W
Minimum pulse width 50us
Current accuracy 0.04%+1mA (Low Range)
0.08%+5mA (High Range)

Note: Product specifications and descriptions in this document are subject to change without notice.